HFM-25熱流儀提供全尺寸熱流量計的測試能力,並針對小件隔熱與建材樣品和預算進行優化與升級。
HFM-25經濟型的設計滿足對小型隔熱樣品真實熱阻和熱導率進行穩態測試日益增長的的需求。
秉承簡約設計的理念,樣品被放置在加熱-冷卻兩塊測試板之間,通過手動控制調節與板的接觸情況,並可用標準砝碼施加可變壓力。它自帶的數字儀表方便提供測量厚度,使用附帶的墊片可對可壓縮的樣品進行測試。
該設備專為室溫操作而設計,用戶可以從 5°C、10°C、15°C 和 20°C 四個增量中進行選擇。每個板內都集成了一個熱流感測器,用於定時監測上測試板和下測試板因溫差(ΔT)產生的熱流(Q/A)。定時監測會一直持續,直至觀察到穩態熱流,然後利用復合熱流測量熱阻(R),並根據傅里葉定律計算熱導率(λ)。
Following international standards, the HFM Series is designed for testing both homogeneous and heterogeneous materials of a range ofsample dimensions.
Methods | HFM-100 | HFM-100 HT | HFM-50 | HFM-25 |
---|---|---|---|---|
Materials | Insulation, solids, and textiles | Insulation, solids, and textiles | Insulation, solids, and textiles | Insulation, solids, and textiles |
Type of Sensors | Flux sensors (x2) | Flux Sensors (x2) | Flux sensors (x2) | Flux sensors (x2) |
Thermal conductivity (W/m•K) | 0.002 to 0.5 | 0.002 to 0.5 | 0.002 to 0.5 | 0.01 to 0.3 |
Specific heat | Optional | Optional | Optional | N/A |
High thermal conductivity kit (W/m•K) | Optional (up to 2.5) | Optional (up to 2.5) | Optional (up to 2.5) | N/A |
Sample size (mm) | 300 x 300 x up to 100 | 300 x 300 x up to 100 | 200 x 200 x up to 50 | Up to 300 x 300 x 25 |
Test time (minutes) | 30 to 40 | 30 to 40 | 30 to 40 | 20 |
Accuracy (Thermal conductivity) * | 1 to 2% | 1 to 2% | 1 to 2% | 3% |
Repeatability (Thermal conductivity) | 0.5 to 1% | 0.5 to 1% | 0.5 to 1% | 1% |
Plate temperature range (°C) | 20 to 75** | -30 to 110** | -20 to 70** | 10 to 75 |
Factory calibrated | Yes | Yes | Yes | Yes |
Standard | ASTM C518, ASTM C1784, ISO 8301, JIS A1412, EN 12667, and EN 12664 | ASTM C518, ASTM C1784, ISO 8301, JIS A1412, EN 12667, and EN 12664 | ASTM C518, ASTM C1784, ISO 8301, JIS A1412, EN 12667, and EN 12664 | ASTM C518, ASTM C1784, ISO 8301, JIS A1412, and EN 12667 |
*Performance verified with NIST 1450d / 1450e
**Chilled circulator required
用HFM-25對NIST 1450d高密度纖維玻璃板的均勻性進行測量。該設備專為測試小尺寸樣品或對大尺寸樣品進行25 mm x 25 mm的多點選點測試而設計。同時,開放式樣品夾可以用於測試大型樣品。
通過對板的四個角進行測量來分析樣品的均勻性。若不同位置的% RSD(相對標準偏差百分比)小於1%,則證明其具有出色的均勻性。
用HFM-25對氣凝膠毯樣品(50 x 50 x 9.58 mm)進行測試。在23℃下測得熱導率為0.014 W/m•K,這在(通過GHP – ASTM C177進行測量的)製造商規格所規定的3%以內。
為了確保準確性,對NIST SRM 1450d樣品的熱導率進行了測量。NIST樣品在25°C的認證熱導率為0.0329 W/m•K。測得的熱導率為0.0330 W/m•K,在認證值誤差的1%以內。